Comparison of novel 6.5 Fr sheathless guiding catheters versus 5 Fr guiding catheters for transradial coronary intervention
Chiam Paul, Liu B, Wong AS, Wong PE, Lim ST, Koh TH, Lim VY.
EuroIntervention 2011; 7: 930-5.
AIMS:
To assess the safety and efficacy of a novel sheathless (SH) 6.5 Fr (French) hydrophilic-coated guiding catheter (GC) compared to the standard 5 Fr GCs in transradial coronary interventions (TRI).
METHODS AND RESULTS:
Patients undergoing TRI with 6.5 Fr SH or 5 Fr GCs were included. Baseline characteristics and in-hospital outcomes were recorded. Primary endpoints were procedural success and presence of radial pulse at discharge. Secondary endpoints were successful GC support, in-hospital adverse events, access-site complications, procedural duration and contrast load. There were 269 patients with 146 procedures in each group. The SH GC group had more non-ST elevation MI, in-stent restenosis, high-risk and bifurcation lesions. Procedural success in both arms was 95.2%. One patient in each group (0.7%) experienced radial artery occlusion (RAO) after TRI, without clinical sequelae. One access-site haematoma and one minor stroke occurred in the 5 Fr group (none in the SH group, both p=ns). Mean procedure time (52±21 vs. 45±21 minutes, p=0.004) and contrast load (160±45 ml vs. 140±45 ml, p=0.003) were greater in the SH group.
CONCLUSIONS:
Both 6.5 Fr SH GCs and 5 Fr GCs achieved high procedural TRI success with low RAO rates. The SH GC eliminated the disadvantages of the 5 Fr GC whilst maintaining the advantage of low RAO rates, and may become the GC of choice in TRI.